IMPEDANCE MODEL OF PRECISION PLATINUM RESISTANCE THERMOMETERS IN THE FIELD OF HIGH TEMPERATURES
ARTICLE_12_PDF (Українська)

Keywords

temperature
measurement
resistance thermometer
frequency error температура
измерение
термометр сопротивления
частотная погрешность

How to Cite

[1]
Михаль, А. and Мелещук , Д. 2012. IMPEDANCE MODEL OF PRECISION PLATINUM RESISTANCE THERMOMETERS IN THE FIELD OF HIGH TEMPERATURES. Tekhnichna Elektrodynamika. 4 (May 2012), 073.

Abstract

Double-pole multiunit impedance model of sensitive thermometer element during measurement on alternating current is proposed. It takes into consideration impact on resistance of sensitive element of bypassing effect which is caused by essential lowing of specific resistance of framework isolation elements with temperature growth. As a result, during the measurement of thermometer resistance on direct and alternating current the error appears which is the function of measurable temperature and current frequency. Theoretical investigations of frequency error of thermometers in the range of temperatures from 0 to 1100°C are conducted. Difference in the results of measurement of resistance of major types of thermometers is established which leads to the error of 0,1-0,2 mK Model analysis demonstrates the necessity for introduction of individual frequency correction to the measurement result like correction for zero power of dispersion. References 10, table 1, figures 9.

ARTICLE_12_PDF (Українська)

References

Glukhenkii A.I., Mikhal A.A. Calculated mark of impedance composed of cylindrical conductor at its measurement on alternating current // Tekhnichna elektrodynamika. - 2011. - № 1. - Pp. 15-22. (Rus)

Zernov N.V., Karpov V.G. Theory of radiotechnical chaihs. - Moskva: Energiia, 1965. - 892 p. (Rus)

Kuin P. Temperature: Transl. from Engl. - Moskva: Mir, 1985. - 448 p. (Rus)

Lutsyk Ya.T., Guk O.P., Lakh O.I., Stadnyk B.I. Temperature measurements: theory and practice. - Lviv: “Bleksyd Bit” edition. - 2006. - 560 p. (Ukr)

Meleshchuk D.V., Mikhal A.A. Error of platinum resistance thermometers, caused by surface effect in a wire of sensitive element // Tekhmchna elektrodynamika. - 2004. - № 2. - Pp. 69-71. (Rus)

Reference book on theoretical basis of radiotechnics. (P.195). By Krivitskii B.Kh., Dulin V.N. - Vol. l. - Moskva: Energiia, 1977. - 504 p. (Rus)

Reference book on electrotechnical materials: (P.74). By Koritskii Yu.V. and others. 3 volumes. Vol. 2 - 3-d edition. - Moskva: Energoatomizdat, 1987. - 464 p. (Rus)

Compton I.P. The realization of low temperature fixed points. TMCSI, 4. - 1972. - Part 1. - Pp. 195-209.

Evans J.P., Burns G.W. A study of stability of high temperature platinum resistance thermometers. TMCSL 3.- 1962.-Part l.-Pp. 313-318.

Long Guang, Tao Hongtu. Stability- of precision hieh temperature platinum resistance thermometers. TMCSI, 5. - 1982. - Part 1. - Pp. 783-787.

Creative Commons License

This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.

Copyright (c) 2023 Array

Abstract views: 33 | PDF Downloads: 5