OPTIMIZATION OF BALANCING IN A BRIDGE MEASURING CIRCUIT WITH A DIFFERENTIAL CONDUCTOMETRIC SENSOR
ARTICLE_13_PDF

Keywords

differential conductometric biosensors
impedance
measurement
common mode influences
equivalent electrical model диференціальні кондуктометричні біосенсори
імпеданс
вимірювання
синфазні впливи
еквівалентна електрична модель

How to Cite

[1]
Melnyk, V., Borshchov, P., Vasylenko, O. and Brahynets, I. 2022. OPTIMIZATION OF BALANCING IN A BRIDGE MEASURING CIRCUIT WITH A DIFFERENTIAL CONDUCTOMETRIC SENSOR. Tekhnichna Elektrodynamika. 2022, 4 (Jul. 2022), 078. DOI:https://doi.org/10.15407/techned2022.04.078.

Abstract

The article is devoted to solving the problem of the occurrence of an additive error in determining local changes in the electrical conductivity of electrolyte solutions under conditions of changes in the background electrical conductivity of the measurement medium, which often occurs in biosensor and other systems with a differential pair of conductometric transducers, if their electrical parameters are not identical. The goal is to provide a deep suppression of the influence of background changes with significant differences in both reactance and active resistance in the transducers of a pair of sensor. The essence of the issue, the causes and mechanism of this type of error, as well as the methods and means of its reduction, developed earlier, are briefly considered. A diagram and description of the structure of a differential conductometric channel of a biosensor system based on an AC bridge, an algorithm for its balancing operations by controlling the module and phase of the test voltage, as well as a vector diagram of currents and voltages in the bridge circuit during this process. The balancing of the bridge has been was modeled analytically, bringing it to a quasi-equilibrium state, in which changes in the background electrical conductivity do not change its output signal. Additional operations for balancing the bridge are determined to achieve such a state with significant differences in both capacitances and active resistances in the impedances of a pair of conductometric transducers of a differential sensor. The results of experimental studies of the suppression of the influence of changes in the background electrical conductivity of a solution in a differential conductometric channel with using its computer model and experimental sample of a conductometric instrument with an electrical equivalent of a differential sensor are presented. A comparison of the results obtained and the corresponding data for balancing bridge circuits by previously developed methods is given. References 16, figures 3, tables 3.

https://doi.org/10.15407/techned2022.04.078
ARTICLE_13_PDF

References

Impedance Spectroscopy Theory, Experiment, and Application. Edited by Evgenij Barsoukov, Ross Macdonald. John Wiley & Sons Inc., Hoboken, New Jersey. 2005.

Grossi M., Riccò B. Electrical impedance spectroscopy (EIS) for biological analysis and food characterization: a review. Journal of Sensors and Sensor Systems. 2017. No 6. Pp. 303-325. DOI: https://doi.org/10.5194/jsss-6-303-2017.

Narjes Kolahchi, Mohamed Braiek, Gholamhossein Ebrahimipoura, Seyed Omid Ranaei-Siadatc, Florence Lagarde Nicole Jaffrezic-Renaultb. Direct detection of phenol using a new bacterial strain-based conductometric biosensor. Journal of Environmental Chemical Engineering. 2018. Vol. 6. Issue 1. Pp. 478-484. DOI: https://doi.org/10.1016/j.jece.2017.12.023

Lee R., Kester W. Fully Automatic Self-Calibrated Conductivity Measurement System. Analog Devices: Analog Dialogue 50-11. 2016.

Jaffrezic-Renault N., Dzyadevych S.V. Conductometric microbiosensors for environmental monitoring. Sensors (Basel). 2008. Vol. 8. Issue 4. Pp. 2569-2588. DOI: https://doi.org/10.3390/s8042569

Dzyadevich S.V., Soldatkin O.P. Scientific and technological principles of creating miniature electrochemical biosensors. Kyiv: Naukova dumka, 2006. 256 p.

Dzyadevych S.V. Conductometric enzyme biosensors: theory, technology, application. Analytica Chimica Acta. 2001. Vol. 445. Pp. 47-55. DOI: https://doi.org/10.1016/S0003-2670(01)01249-1

Grinevich F.B., Surdu M.N. High-precision variation measuring systems of alternating current. Kyiv: Naukova Dumka, 1989. 192 p. (Rus)

Surdu M.M., Monastirsky Z.Ya. Variations of methods for improving the accuracy of vimiruvachiv imitation. Kyiv, Institute of Electrodynamics of the National Academy of Sciences of Ukraine, 2015. 385 p. (Ukr)

Melnyk V.G., Rubanchuk M.P., Mikhal A.A. Measuring circuits for conductometric transducers with differential two-electrode sensors. Tekhnichna electrodynamika. 2008. No 2. Pp. 119-124. (Rus)

Melnyk V.G., Dzyadevych S.V., Ivashchuk A.V., Ulyanova V.A., Lepikh Ya.I., Romanov V.O. The experimental studies of microelectronic transducers for conductometric biosensor systems. Sensorna elektronika ta mikrosystemni tekhnolohii. 2011. Vol. 8. Issue 3. Pp. 81–90. DOI: https://doi.org/10.18524/1815-7459.2011.3.118131 (Rus)

Melnyk V.G., Vasylenko A.D., Semenycheva L.N., Slitskiy O.V., Saiapina O.Y., Dzyadevych S.V. Solutions for enhancement of sensitivity and metrological reliability of conductometric biosensor systems. Engineering Research Express. 2021. Vol. 3. No 4. DOI: https://doi.org/10.1088/2631-8695/ac2a0d

Melnik V.G., Vasilenko A.D., Dudchenko A.E., Pogrebnyak V.D. Studies of common-mode interference suppression in a biosensor conductometric system with differential sensors. Sensorna elektronika ta mikrosystemni tekhnolohii. 2014. Vol. 11. No 3. Pp. 49-61. URL: http://semst.onu.edu.ua/article/view/108258 (accessed at 12.03.2022). (Rus) DOI: https://doi.org/10.18524/1815-7459.2014.3.108258

Dudchenko O.E., Matsyshyn M.Y., Peshkova V.M., Soldatkin O.O., Soldatkin O.P., Dzyadevich S.V. Methods of testing conductometric transducers for further biosensor use. Sensorna elektronika ta mikrosystemni tekhnolohii. 2013 Vol. 10. No 4. Pp. 97-109. (Ukr)

Melnyk V.G., Onyshchenko I.V., Rubanchuk M.P., Slitsky A.V. Improving common mode interference suppression in a differential conductometric biosensor system. Tekhnichna electrodynamika. 2015. No 2. Pp. 73-82. (Rus)

Melnyk V.G., Borshchov P.I., Dzyadevych S.V., Saiapina O.Y., Vasylenko O.D. Increasing the sensitivity and metrological reliability of a differential conductometric biosensor system. Tekhnichna electrodynamika. 2021. No 6. Pp. 68-78. DOI: https://doi.org/10.15407/techned2021.06.068.

Creative Commons License

This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.

Copyright (c) 2022 Array

Abstract views: 258 | PDF Downloads: 124

Downloads