Abstract
The mathematical model for error estimation of the conversion of standard four-electrode conductometric cell conditional on non-equipotentiality of electrode face has been worked out. The conditions under which this error does not exceed 10 ppm have been specified. References 6, figures 5.
References
Gavrylkin V.G. ta in. The state primary standard of unit of electrolytic conductivity of liquids // Ukrainskyi metrologichnyi zhurnal. 2006. No3. P. 47-51. (Ukr.)
Glukhenkii A.I., Mykhal A.A. Analysis of circular conductor impedance component within AC measurement // Tekhnichna elektrodynamika. 2010. No 1. P. 15-22. (Rus.)
Karandeev K.B., Shtamberger G.A. The generalized theory of AC bridge circuits. Novosibirsk: Izdatelstvo Sibirskogo Otdeleniia Akademii Nauk SSSR, 1961. 224 p. (Rus.)
Brinkmann F., Ebbe Dam N., Deák E., Durbiano F., Ferrara E., Fükö J., Jensen H. D., Máriássy M., Shreiner R. H., Spitzer P., Sudmeier U., Surdu M., Vyskocil L. General paper: Primary methods for the measurement of electrolytic conductivity // Accred Qual Assur. – 2003. – 8:346 – 353 DOI 10.1007/s00769-003-0645-5.
Wu Y.C. A DC method for the absolute determination of conductivities of the primary standard KCl solution from 00C to 500C // J. Res. Nati. Inst. Stand. Technol. 1994. Vol. 99. No 3. P. 241-246.
http://kcdb.bipm.org/AppendixB/appbresults/ccqm-k36/ccqm-k36_final_report.pdf
This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.
Copyright (c) 2023 Array